MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® NO ADDITIVE (Z) TUBES; SPECIMEN TRANSPORT AND STORAGE CONTAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® NO ADDITIVE (Z) TUBES; SPECIMEN TRANSPORT AND STORAGE CONTAINER Back to Search Results
Model Number 366703
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® no additive (z) tube there was foreign matter in the tube(s) biological and non-biological.This event affected 29 devices.The following information was provided by the initial reporter.The customer stated: "black spots were found on several tubes.".
 
Additional Manufacturer Narrative
H6: investigation summary.Bd received twenty-nine (29) samples for investigation.The samples were evaluated by visual examination and the indicated failure mode for foreign matter (black spots) was observed embedded in the sidewall of the tubes with the incident lot.Embedded foreign matter is a cosmetic defect, as it is isolated from any specimen.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality notifications.This complaint has been confirmed for the indicated failure mode of embedded foreign matter (black spots) based on the returned samples.Bd determined that the root cause of the indicated failure mode was attributed to manufacturing.Based on an evaluation of severity and frequency it was determined that no corrective action is required at this time.
 
Event or Problem Description
It was reported when using the bd vacutainer® no additive (z) tube there was foreign matter in the tube(s) biological and non-biological.This event affected 29 devices.The following information was provided by the initial reporter.The customer stated: "black spots were found on several tubes.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® NO ADDITIVE (Z) TUBES
Common Device Name
SPECIMEN TRANSPORT AND STORAGE CONTAINER
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13905739
Report Number1917413-2022-00155
Device Sequence Number14601472
Product Code FMH
UDI-Device Identifier50382903667036
UDI-Public50382903667036
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2022
Device Model Number366703
Device Catalogue Number366703
Device Lot Number1040913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2022
Supplement Date Received by Manufacturer04/01/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-