MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL Back to Search Results
Model Number BXA085902A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.A review of the manufacturing records indicated the device met pre-release specifications.Due to no device return, an investigation could not be performed.
 
Event or Problem Description
It was reported to gore that during a procedure for treatment of the external iliac and common iliac, the gore® viabahn® vbx balloon expandable endoprosthesis came detached from the balloon while inside the patient.Reportedly the physician was unable to cross the lesion of the common iliac and attempted to draw the device back into the sheath, at which point the device came detached from the balloon.It was reported that the device was able to be implanted at the second target site in the external iliac.Reportedly the procedure to treat the common iliac was completed using another gore® viabahn® vbx balloon expandable endoprosthesis.It was reported that there was no adverse patient outcome and the patient tolerated the procedure.
 
Additional Manufacturer Narrative
Correction: this event was a product malfunction.Addition of investigation conclusion code 61.Product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Common Device Name
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
spencer deboard
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13906124
Report Number2017233-2022-02824
Device Sequence Number14522304
Product Code PRL
UDI-Device Identifier00733132637331
UDI-Public00733132637331
Combination Product (Y/N)Y
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P160021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2024
Device Model NumberBXA085902A
Device Catalogue NumberBXA085902A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer03/07/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient RaceBlack Or African American
-
-