• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A30 VENTILATOR (A SERIES, MERLIN); VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A30 VENTILATOR (A SERIES, MERLIN); VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number 1116151
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2016
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned for service.There was no harm or injury reported.During the evaluation at the manufacturer's service center, the sound abatement foam was replaced for repair.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
 
Additional Manufacturer Narrative
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
 
Additional Manufacturer Narrative
The manufacturer previously reported life-threatening in b.2 section, which was marked incorrectly as this is product problem only as mentioned in b.1 section in previous report.G.6 section incorrectly marked 30 days and initial including 15 days, which should marked as 30 days and initial only in this follow up report.G.2 section incorrectly marked in previous report, which is correctly marked as consumer and foreign in this follow up report.E.1 section was missing in previous report, which is correctly updated in this follow up report.D.8 was incorrectly reported in the previous report, which is correctly updated in this follow up report.Section h.3 was marked incorrectly in previous report, which is correctly reported in this follow up report.Type of investigation in h.6 section was incorrectly coded as 4105 and 10, which is correctly updated as 4114 in this follow up report.Udi number in d.4 section was missing in the previous report, which is correctly update now.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A30 VENTILATOR (A SERIES, MERLIN)
Common Device Name
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key13909018
Report Number2518422-2022-11379
Device Sequence Number11859778
Product Code MNS
UDI-Device Identifier00606959042827
UDI-Public00606959042827
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K113053
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1116151
Device Catalogue Number1116151
Was Device Available for Evaluation? No
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 03/22/2016
Supplement Date Received by Manufacturer03/22/2016
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date04/18/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/18/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES 88058
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-