Model Number 9-ASD-030
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The additional amplatzer septal occluder lot 7170020 is filed under separate manufacturer reference number: 2135147-2022-00104-00.
 
Event or Problem Description
It was reported, a 30mm amplatzer septal occluder (7023673) was chosen for procedure with a 12f amplatzer trevisio intravascular delivery system.During procedure, upon deployment of the device, the left disc of the device presented in a cobra shape deformation.The device was retrieved and a new, 32mm amplatzer septal occluder (7170020) was deployed using the same delivery system.This device also presented with a cobra shape deformation of the left disc.The device was re-sheathed twice.The device was removed and exchanged with a 34 mm amplatzer septal occluder (7232844).The 34mm amplatzer septal occluder was determined a mis-sizing, the device was too big therefore never implanted.The patient remained hemodynamically stable throughout the procedure.Ultimately no device was implanted.
 
Additional Manufacturer Narrative
An event of device deformity was reported.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.No indication of a lot related issue was found.The cause of the reported event could not be conclusively determined.Please note, per the instructions for use, states the minimum recommended size delivery system for use with a 30 mm amplatzer septal occluder is an 10f.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13911981
Report Number2135147-2022-00105
Device Sequence Number14522183
Product Code MLV
UDI-Device Identifier05415067019332
UDI-Public05415067019332
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model Number9-ASD-030
Device Catalogue Number9-ASD-030
Device Lot Number7023673
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/02/2022
Supplement Date Received by Manufacturer04/20/2022
Initial Report FDA Received Date03/25/2022
Supplement Report FDA Received Date05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMPLATZER TREVISIO DS, LOT 7500550
Patient SexFemale
-
-