| Model Number |
VERSION 1.0 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.(expiry date: 08/2024).
|
| |
|
Event or Problem Description
|
|
It was reported that during a recanalization procedure, the helix was allegedly found to be broken inside the vessel.The broken helix was completely removed from the body without any issues.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.Manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned for evaluation.A physical investigation was performed for the catheter.The helix was broken at 15 cm distal.It can be confirmed that the helix was broken.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiry date: 08/2024).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a recanalization procedure, the helix was allegedly found to be broken inside the vessel.The broken helix was completely removed from the body without any issues.There was no reported patient injury.
|
| |
|