• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L331
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885); High Capture Threshold (3266)
Patient Problem Atrial Fibrillation (1729)
Event Date 10/01/2021
Event Type  malfunction  
Event Description
It was reported during a routine office visit, that this pacemaker device was suspected of behaving in a manner consistent with a premature battery depletion (pbd).A technical service (ts) consultant reviewed data with the physician, noting that the device battery longevity had eight years remaining in november 2020, and now the battery longevity is six months remaining.Ts discussed that the patient is in chronic atrial fibrillation since october 2021, (100% of every day).Thresholds have been increasing from 1.1v to 1.6v.Ts provided recommendations for programming to help extend battery longevity.Ts confirmed this device does not have a premature battery depletion.The physician will recheck the patient in 3 months.The device remains implanted.No adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited sensing of chronic high atrial rates which may impact battery longevity over time.Chronic high atrial sensing rates, such as for patients with chronic atrial fibrillation, can reduce longevity in devices that are programmed with atrial sensing on.Device power consumption increases proportional to the additional processing for sensed atrial events.As a result, the high rate of atrial sensing can result in the longevity being lower than expected, as observed with this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI EL DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13912439
MDR Text Key288054270
Report Number2124215-2022-06225
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/06/2022
Device Model NumberL331
Device Catalogue NumberL331
Device Lot Number882927
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
-
-