Device evaluation summary: upon visual evaluation of the image provided in the complaint, no apparent damage or visual anomalies were observed.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.A manufacturing record evaluation was performed for the finished device 7283936 number, and no non-conformances were identified.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Reason for device explant and/or reoperation: capsular contracture baker grade iii manufacturer¿s reference number: (b)(4).
|
|
On april 6, 2022 the mentor failure analysis lab has received the device for evaluation.Patient's initials are (b)(6).Date of event is (b)(6) 2021.Date of explantation is (b)(6) 2021.The investigation of the returned device was completed by the failure analysis lab on april 26, 2022.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During the visual evaluation, no apparent damage or visual anomalies were observed on the sm mod classic gel, 340cc returned device.Infection, manifested by swelling, tenderness, pain, and fever, may appear in the immediate postoperative period or at any time after insertion of the implant.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If the infection does not subside promptly with the appropriate treatment, removal of the implant is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|