MAUDE Adverse Event Report: TELEFLEX MEDICAL HZ APPLIER MED 8" CVD; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HZ APPLIER MED 8" CVD; APPLIER, SURGICAL, CLIP Back to Search Results
Model Number IPN028188
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
Received call from senior staff nurse to be advised that during a coronary artery bypass graft they had a number of clip failures (10+).The nurse was not in the case so she is unaware of how it was resolved.She believes that the patient's condition is okay.The nurse is going to confirm the lot number of the clips used and is requesting for the appliers to be decontaminated and quarantined.I have arranged to go in wednesday to inspect the appliers and gather further details from surgical team.I am advised photos were taken and the clinical lead was called in to observe the issue.The appliers were new appliers.They have been used only during the past 3 week.I will be able to identify how many times they have been processed after wednesday.Clinical consequences: medical intervention was necessary due to the alleged issue.Additional information: it is probable that sutures were needed to ligate the vessel properly but she is waiting for confirmation from the surgeon.She collected 5 cartridges of representative clips and the appliers.All four of the medium appliers would have been used during the procedure but she could not say which one if any was responsible for the problem.She inspected them visually and could not observe any default.The small appliers were collected by the sales representative but they are not relevant to this complaint.
 
Additional Manufacturer Narrative
(b)(4).The device investigation is pending.Teleflex will continue to monitor and trend related events.
 
Additional Manufacturer Narrative
(b)(4).The dhr for the returned device was reviewed and found completely without any irregularities.This device was produced at the tecomet, inc.Kenosha wi facility as part of a 100 piece lot in august of 2020.Evaluation of the returned instrument shows that the tips are aligned in the open and closed position and that this instrument is able to pick-up, retain, closed and release multiple clips both with and without the use of silastic test tubing.Further evaluation using this instrument to close 5 clips shows that maximum eye gap specification of.0073 measured at (.0019-.0034 max.) and the maximum leg gap specification of.0060 measured at (.0035-.0042 max.) verifying that this instrument performs as designed.We were unable to find any nonconforming features on this device therefore we are unable to validate this complaint.We are unable to determine what caused the alleged defect at the end user's facility.All 100 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Event or Problem Description
Received call from senior staff nurse to be advised that during a coronary artery bypass graft they had a number of clip failures (10+).The nurse was not in the case so she is unaware of how it was resolved.She believes that the patient's condition is okay.The nurse is going to confirm the lot number of the clips used and is requesting for the appliers to be decontaminated and quarantined.I have arranged to go in wednesday to inspect the appliers and gather further details from surgical team.I am advised photos were taken and the clinical lead was called in to observe the issue.The appliers were new appliers.They have been used only during the past 3 week.I will be able to identify how many times they have been processed after wednesday.Clinical consequences: medical intervention was necessary due to the alleged issue.Additional information: it is probable that sutures were needed to ligate the vessel properly but she is waiting for confirmation from the surgeon.She collected 5 cartridges of representative clips and the appliers.All four of the medium appliers would have been used during the procedure but she could not say which one if any was responsible for the problem.She inspected them visually and could not observe any default.The small appliers were collected by the sales representative but they are not relevant to this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HZ APPLIER MED 8" CVD
Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key13915905
Report Number3011137372-2022-00038
Device Sequence Number3883473
Product Code GDO
UDI-Device Identifier24026704710038
UDI-Public24026704710038
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN028188
Device Catalogue Number237081
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Supplement Date Received by Manufacturer04/12/2022
Initial Report FDA Received Date03/26/2022
Supplement Report FDA Received Date04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-