• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL HEAL COLLAR 3.5X4.5, 5MM; ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL HEAL COLLAR 3.5X4.5, 5MM; ABUTMENT Back to Search Results
Model Number HC345
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the healing collar would not seat due to the threads being off.Procedure was completed using another healing collar.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Patient identifier: patient id not provided.Age or date of birth: patient age not provided.Weight: patient weight not provided.
 
Additional Manufacturer Narrative
One (1) heal collar 3.5x4.5, 5mm (hc345) was returned for investigation.Visual evaluation of the as returned product identified no obvious signs of deviations.However, when tried into stock tsv the threads would not engage.Functional testing was not performed due to the nature of the device and event.The product could not seat.Therefore, the reported event was recreated.A pre-existing condition noted on the per was unknown bone density type.The reported device was located on an unknown tooth site (universal).The customer did not provide any pictures or x-rays.Review of appropriate documentation: documents reviewed: instructions for use ¿ zimmer dental healing collars, healing screws, surgical cover screws, temporary gingival cuffs, and temporary abutments for use with zimmer dental implant systems, ifu9658 rev 2-10/19.Information identified: indications and warnings per the applicable ifu, improper techniques and overloading may lead to device failure.Dhr review: dhr review was completed for the subject lot number (2020070136).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (2020070136) for similar events and one (1) other complaint was identified (b)(6).Review completed utilizing keywords: does not seat post market trending review: july post market trending was reviewed and there were no actionable events or corrective actions for the reported event (does not seat) or product (hc345).No actionable items have been triggered that will affect complaint handling on our end for this month.Therefore, based on the available information, device malfunction did occur and the reported event was confirmed.Sections updated: b4: date of this report.G3: date investigation results were received.G6: type of report and follow up number.H2: follow up type.H3: device evaluated.H6: adverse event problem codes.H10: manufacturer's narrative.
 
Event or Problem Description
There is no update to the original complaint description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEAL COLLAR 3.5X4.5, 5MM
Common Device Name
ABUTMENT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key13919334
Report Number0002023141-2022-00722
Device Sequence Number5876857
Product Code NHA
UDI-Device Identifier00889024015005
UDI-Public(01)00889024015005(17)250713(10)2020070136
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K111852
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/13/2025
Device Model NumberHC345
Device Catalogue NumberHC345
Device Lot Number2020070136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date03/26/2022
Supplement Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-