| Model Number |
HC345 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Patient id: patient id not provided.Age or date of birth: patient age not provided, weight: patient weight not provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the threading was off and the abutment was unable to be seated.The procedure was completed using another device.
|
| |
|
Event or Problem Description
|
|
There is no update to the original complaint description provided.
|
| |
|
Additional Manufacturer Narrative
|
|
A heal collar 3.5x4.5, 5mm (hc345) was returned for investigation.Visual inspection of the as returned product identified worn markings due to usage and damaged internal threads identified functional testing was performed for the returned product with an in-house stock implant and engages and disengages as intended however with force due to the damage threads.Pre-existing patient factors, x-ray, tooth location are not relevant to the reported event.The length of the device usage is unknown.Pictures images were not provided.Review of appropriate documentation: documents reviewed: instructions for use for healing collars, healing screws, surgical cover screws and temporary gingival cuffs 9658 rev 2 10/19.Information identified: description and adverse effects.Dhr review: dhr review was completed for the subject lot number (2021020071).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (2021020071) for similar event and no other complaint was identified.Review completed utilizing keywords: damaged threads.Post market trend review: july post market trending was reviewed and there were no actionable events or corrective actions for the reported event or device.Based on the available information, device malfunction did occur and the reported event was confirmed.
|
| |
|
Search Alerts/Recalls
|