Model Number LIMB-O¿ SINGLE LIMB ANESTHESIA BREATHING CIRCUIT
Medical Device Problem Code Failure to Disconnect (2541)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On the investigation, since no sample or pictures were provided we were unable to confirm the reported defect.Sample or pictures are needed to perform a better investigation.Therefore, the root cause was not determined.In addition pfmea 57a was reviewed and we have controls for this kind of issues.Any additional information provided by the customer will be included in a follow up report.
 
Event or Problem Description
The customer reported an incident with afn52004 limb-o¿ single limb anesthesia breathing circuit.The anesthetist running flows at 10 liters initially with adjustments as high as 16 liters.Running 2.3 sevoflurane during whole case.Anesthetist turned gas off early to finish case.Duration of sevoflurane gas off from 535 to 550 before being able to extubate at 553.Exp sevoflurane gas read at 0.2 and unable to reduce in timely fashion.Corrective measures taken included flushing circuit with fresh gas, still without timely reduction in exp sevoflurane.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIMB-O¿ SINGLE LIMB ANESTHESIA BREATHING CIRCUIT
Common Device Name
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
PRODUCTOS UROLOGOS DE MEXICO S.A. DE C.V.
cerrada via de la produccion
no. 85, parque undustrial mex
mexicali, 21397
MX   21397
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key13925185
Report Number8030673-2022-00238
Device Sequence Number1778981
Product Code CAI
UDI-Device Identifier10190752121775
UDI-Public(01)10190752121775
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLIMB-O¿ SINGLE LIMB ANESTHESIA BREATHING CIRCUIT
Device Catalogue NumberAFN52004
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Initial Report FDA Received Date03/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-