Name and address: phone: (b)(6).Device evaluated by mfr: device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
It was initially reported, during ureteroscopic lithotripsy (ursl), when the package of a universa soft ureteral stent set was opened it was discovered that one side of the pigtail loop had a "rupture".Another same type device was used to complete the procedure without difficulties.The stent was returned to the manufacturer, and inspection of the returned device found the the stent was torn, almost entirely around the circumference of the stent.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.
|
|
Event summary: it was initially reported, during ureteroscopic lithotripsy (ursl), when the package of a universa soft ureteral stent set was opened it was discovered that one side of the pigtail loop had a "rupture".Another same type device was used to complete the procedure without difficulties.The stent was returned to the manufacturer, and inspection of the returned device found the the stent was torn, almost entirely around the circumference of the stent.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.Investigation ¿ evaluation: a visual inspection of the returned device was conducted.A document-based investigation was also performed including a review of complaint history, device history record (dhr), the instructions for use, specifications, interview with personnel and quality control data.A device failure analysis was conducted: one device was returned in open packaging.Tether is included, but not attached to stent.First distal side port has been torn, almost entirely around the circumference of the stent.There is another tear approx.4.1cm from the distal end, near the most proximal side port.The conclusion of the dfa was that the stent was broken on the untethered side; however, the situation or cause(s) leading up to the break is unknown.A review of the device history record (dhr) found no non-conformances related to the reported failure mode.A review of complaint history records shows no other complaints associated with the complaint device lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: how supplied: upon removal from package, inspect the product to ensure no damage has occurred.The complaint, reported by the customer of the stent being broken at the tip, was confirmed due to customer testimony and examination by cook.Per the quality engineering risk assessment, no further action is warranted.Cook medical will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|