DEPUY ORTHOPAEDICS INC US TAPER CEM STEM 13X90; UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS Back to Search Results
Model Number 86-6468
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Pain (1994)
Date of Event 10/13/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon reported to (b)(6): "sudden increase of pain in the left knee joint after a fall event.".
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: visual examination of the returned cemented stem found it fractured, also absence of cement and burnishing was found, which is indicative of loosening of the device in situ.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary device history lot: 1) quantity manufactured: (b)(4).2) date of manufacture: 01/20/17.3) any anomalies or deviations identified in dhr: none.4) expiry date: 12/31/26.5) ifu reference: (b)(4).Device history review: 1) quantity manufactured: (b)(4).2) date of manufacture: 01/20/17.3) any anomalies or deviations identified in dhr: none.4) expiry date: 12/31/26.5) ifu reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAPER CEM STEM 13X90
Common Device Name
UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS 1818910
700 orthopaedic dr
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key13928010
Report Number1818910-2022-05582
Device Sequence Number13897936
Product Code JWH
UDI-Device Identifier10603295217176
UDI-Public10603295217176
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K882234
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number86-6468
Device Catalogue Number866468
Device Lot NumberD17010782
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LPS DISTAL FEM COMP XSM LT (EX PKT); LPS UNIV TIB HIN INS XSM 12MM; UNIV FEM SLEEVE CEM 20MM; UNK KNEE FEMORAL ADAPTOR LPS; UNK KNEE FEMORAL AUGMENT LPS; UNK KNEE LPS
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-