• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVAILMED S.A.DE C.V.9 INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVAILMED S.A.DE C.V.9 INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 306001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a trial patient who was using an external neurostimulator ens for unknown indications for use.It was reported that there were no patient issues.When provider opened second lead kit for be, lead was unusable.Tip was exposed and frayed at the end, not rigid at all (floppy).Rep provided be lead from their trunk stock and submitted paperwork to give at no charge.Replacement of defective item.The issue was confirmed resolved.No patient/symptoms reported.
 
Additional Manufacturer Narrative
H3 product analysis #704934043:analysis information pli# 10 product id# 306001 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis identified the {x} electrode at the distal end of the lead was bent.Continuation of d10: serial# unknown: product type accessory medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
No new information for event description.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
AVAILMED S.A.DE C.V.9
c.industrial lt.001 mz.105
tijuana,bc 22444
MX  22444
Manufacturer (Section G)
AVAILMED S.A.DE C.V.9
c.industrial lt.001 mz.105
tijuana,bc 22444
MX   22444
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13928022
Report Number3010291427-2022-00011
Device Sequence Number16589254
Product Code EZW
UDI-Device Identifier00763000114077
UDI-Public00763000114077
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/02/2023
Device Model Number306001
Device Catalogue Number306001
Device Lot Number60317934
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer06/06/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date06/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10....".
Patient Age83 YR
Patient SexMale
-
-