• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. SIMON NITINOL FILTER SYSTEM; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. SIMON NITINOL FILTER SYSTEM; VENA CAVA FILTER Back to Search Results
Model Number 2120F
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Extravasation (1842)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing review: a device history record review could not be performed as the lot number is unknown.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Post deployment, fractures and umbrella section with large circumferential impression on the inferior vena cava wall were noted on an unspecified date.The patient had a chronic fractured indwelling inferior vena cava filter.Therefore, the investigation is confirmed for filter limb detachment.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.
 
Event or Problem Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism.At some time post filter deployment, it was alleged that the filter strut detached and the patient had extravasation.Attempts were made to retrieve the filter, however removal of the filter was not confirmed.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMON NITINOL FILTER SYSTEM
Common Device Name
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key13928301
Report Number2020394-2022-90134
Device Sequence Number6356749
Product Code DTK
UDI-Device Identifier00801741040900
UDI-Public(01)00801741040900
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K970099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2000
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2120F
Device Catalogue Number2120F
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/01/2022
Initial Report FDA Received Date03/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-