• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 12MM/130 DEG TI CANN TFNA 360MM/LEFT-STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 12MM/130 DEG TI CANN TFNA 360MM/LEFT-STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 04.037.257S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, the patient underwent an orthopedic procedure.The surgeon was operating a patient and was inserting a left femur nail, while inserting the helical blade, it was determined that the alignment was off.The nail and blade were removed and checked with the aiming arm and insertion handle on back table.It was determined that the insertion handle was bent.Another set was used to complete the surgery successfully.There was a surgical delay of ten (10) minutes.There was no patient consequence reported.This report involves one (1) tfna nail.This is report 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Therapy date: (b)(6) 2022.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h6: a device history record (dhr) review was conducted: manufacturing location: monument.Manufacturing date: 18-jan-2022.Expiration date: 31-dec-2031.Part number: 04.037.257s, 12mm/130 deg ti cann tfna 360mm/left- sterile.Lot number: 603p736 (sterile).Lot quantity:(b)(4).This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12MM/130 DEG TI CANN TFNA 360MM/LEFT-STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13928401
Report Number2939274-2022-01086
Device Sequence Number6251195
Product Code HSB
UDI-Device Identifier10886982097378
UDI-Public(01)10886982097378
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.037.257S
Device Catalogue Number04.037.257S
Device Lot Number603P736
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/26/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
130 DEG AIMING ARM; AIMING ARM LOCKING DEVICE; BLADE/SCREW GUIDE SLEEVE; BUTTRESS/COMPRESSION NUT; COMPLETE RADIOLUCENT INSERTION HANDLE; UNK - NAIL HEAD ELEMENTS: TFNA HELICAL BLADE
Patient Age58 YR
Patient SexFemale
-
-