• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 4.75MM PEEK KNOTLESS SWIVELOCK; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 4.75MM PEEK KNOTLESS SWIVELOCK; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 4.75MM PEEK KNOTLESS SWIVELOCK
Health Effect - Clinical Code Pain (1994)
Date of Event 12/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
On 03/10/2022, it was reported by a sales representative via email that an ar-2324kpslc peek knotless swivelock seems to be fraying and cutting an ar-7535 fiberlink suture tape suture when the suture comes in contact with the metal driver during tensioning.This was discovered during a revision subpec tenodesis procedure on (b)(6) 2022.This issue has caused the bicep to retract and causing patient pain.The original procedure was a biceps loop n tak.Surgeon was able to correct issue with a 2.6 knotless fibertak subpc onlay.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint is unable to be confirmed through the customer provided photos and without receipt of the device for physical evaluation and testing.No change in harm was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.75MM PEEK KNOTLESS SWIVELOCK
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13928473
Report Number1220246-2022-04657
Device Sequence Number6251220
Product Code MAI
UDI-Device Identifier10888867351339
UDI-Public10888867351339
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K193503
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number4.75MM PEEK KNOTLESS SWIVELOCK
Device Catalogue NumberAR-2324KPSLC
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer03/10/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date04/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-