ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012455-12
Medical Device Problem Codes Deflation Problem (1149); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a calcified vessel in the left anterior descending (lad) artery.There was no difficulty removing the stylet or protective sheath of a 5.0x12 mm nc trek rx balloon dilatation catheter (bdc) and the device was prepared (air aspiration) outside the anatomy prior to use.Contrast mix was 50/50 contrast/saline.The balloon was inflated once at 18 atmospheres (atms) and negative was held 30-45 seconds.However, the balloon did not deflate and contrast media remained inside.The balloon could not be pulled back through the guiding catheter.The guiding catheter and guide wires were removed.The guide wires were removed from the guiding catheter and then the physician managed to get the balloon out ultimately deflated.Another same size nc trek bdc was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual, functional and scanning electron microscopy inspections/analysis were performed on the returned device.The reported deflation was confirmed; however, the reported difficulty removing the device could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported deflation issue; however, the reported difficulty removing the device appears to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13928551
Report Number2024168-2022-03193
Device Sequence Number1790242
Product Code LOX
UDI-Device Identifier08717648152146
UDI-Public08717648152146
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number1012455-12
Device Catalogue Number1012455-12
Device Lot Number11117G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/04/2022
Supplement Date Received by Manufacturer05/18/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-