ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1070-120
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 03/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat an unspecified arterial lesion in the lower limb.The 7x120 mm armada 35 percutaneous transluminal angioplasty (pta) catheter was inflated to nominal pressure and the balloon ruptured.It was not possible to remove the balloon without removing the introducer sheath and then the balloon ruptured [separated].The balloon was retrieved in an unspecified manner and a large introducer sheath was positioned to reduce the abundant bleeding and to control the right femoral access.A manual compression was applied.There was a large bleeding and it was not confirmed if a blood transfusion was required.The patient will be followed.No additional information was provided.
 
Additional Manufacturer Narrative
Visual inspection was performed on the returned device.The reported balloon rupture and separation were confirmed.The reported difficulty removing the device could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported balloon rupture.However, balloon material ruptures can be affected by numerous factors including, but not limited to, balloon damage during processing of the balloon material, materials, inflation technique, and interactions with other devices or lesion calcification.The investigation determined the reported difficulty removing the device from the introducer sheath, balloon separation, unexpected medical intervention, removal of foreign body and hospitalization appear to be related to circumstances of the procedure.In addition, the resistance noted during removal and separation of the balloon was likely the result of the ruptured balloon material catching on the anatomy and/or accessory device during removal.A conclusive cause for the reported patient effect of hemorrhage and the relationship to the device, if any, cannot be determined.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13928762
Report Number2024168-2022-03197
Device Sequence Number6251721
Product Code LIT
UDI-Device Identifier08717648154829
UDI-Public08717648154829
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberB1070-120
Device Catalogue NumberB1070-120
Device Lot Number10409G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer05/03/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-