SYNTHES GMBH COLIBRI II HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.101 |
Device Problems
Noise, Audible (3273); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/19/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during an unspecified surgical procedure, it was observed that the colibri handpiece device trigger was not working well (back and forth) and sounded too loud during the operation.It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi:(b)(4).
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.D4: the device serial/lot number was unknown in the initial report and has been updated accordingly.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During evaluation it was determined that the reported condition of the trigger was not working well and was too loud was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device would not run - electrical control unit damage and failed pre-test for check function of the device and check power with the test bench due to component wear.
|
|
Search Alerts/Recalls
|
|
|