The complaint is confirmed.(1) unpackaged ar-1204as-105 was received for evaluation.Visual inspection revealed the 10.5 mm cutter tip, component c2363-10, and the cutter tip pin, component c2364-01 are missing.The distal portion of the actuator tube, component c2363-02, was broken off, and the two protruding features of the distal end of the shaft, component c2361-02, have been twisted and broken.Remnants of tissue were present in the sample.Without additional information, the cause remains undetermined.A probable cause could be attributed to impact with another device or hard bone during the procedure.
|