• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX BILIARY; BILIARY STENT SYSTEM FOR BENIGN STRICTURES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX BILIARY; BILIARY STENT SYSTEM FOR BENIGN STRICTURES Back to Search Results
Model Number M00570380
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Failure to Advance (2524); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a wallflex biliary rx fully covered rmv stent was to be implanted to treat a 4mm malignant biliary stricture during an endoscopic retrograde cholangiopancreatography (ercp) with stent placement procedure performed on (b)(6) 2022.The patient's anatomy was not tortuous and was not dilated prior to stent placement procedure.During the procedure, there was difficulty crossing the delivery system through the stricture.Furthermore, the stent was difficult to deploy and a high amount of force was applied to deploy the stent.The stent was deployed in an incorrect location and was removed from the patient with rat-tooth forceps.Another wallflex biliary stent was implanted to complete the procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX BILIARY
Common Device Name
BILIARY STENT SYSTEM FOR BENIGN STRICTURES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13929174
Report Number3005099803-2022-01542
Device Sequence Number17724075
Product Code PNB
UDI-Device Identifier08714729953197
UDI-Public08714729953197
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
DEN150040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2023
Device Model NumberM00570380
Device Catalogue Number7038
Device Lot Number0027521711
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2022
Initial Report FDA Received Date03/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-