COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number E2515H
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Burn(s) (1757); Full thickness (Third Degree) Burn (2696)
Date of Event 03/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, during procedure, the pencil malfunctioned and burn the side of the mouth.Patient inadvertently burned from the metal that was exposed on electrode where it plugs into the bovie pencil.Electrode and generator used along with the pencil were non-medtronic products.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
New information has been received pertaining to the event.This event has been reassessed and the reportability has been determined to be a serious injury.Additional information: evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that use of the device resulted in a burn.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during an ent (ear, nose, throat) procedure, the pencil malfunctioned and burn the side of the patient's mouth.Patient inadvertently burned from the metal that was exposed on electrode where it plugs into the bovie pencil.The affected area was first treated with a saline soaked gauze then with a silvadene cream.A photograph was received and showed the patient's burn on the mouth was blackened and charred.The procedure was not extended by 30 minutes.Electrode and generator used along with the pencil were non-medtronic products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
tracy landers
5920 longbow drive
boulder, CO 80301
3035816943
MDR Report Key13929506
Report Number1717344-2022-00354
Device Sequence Number1763691
Product Code GEI
UDI-Device Identifier10884524000046
UDI-Public10884524000046
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K813071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberE2515H
Device Catalogue NumberE2515H
Device Lot Number212980354R
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer04/11/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age6 YR
Patient SexMale
Patient Weight21 KG
-
-