Model Number 980X3ENDIUU-RMF
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary medtronic conducted an investigation based upon all information received.It was reported that, while in use on a patient, this 980 ventilator ¿failed on patient¿ and generated a ¿ventilator inoperative¿ message.A review of the diagnostic logs indicates the device entered backup ventilation.The device was available for evaluation.The service personnel (sp) inspected the ventilator and could not confirm the reported issue.Review of the logs indicated the device had entered backup ventilation.Sp checked the unit and found a depleted battery as an additional issue.The ventilator passed all testing per manufacturer specifications at the time of service and was returned to the customer.The investigation could not determine a cause of the event.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that, while in use on a patient, this 980 ventilator "failed on patient" and generated a "ventilator inoperative¿ message.The patient was removed from the ventilator and placed on an alternate ventilator with no injury.A review of the diagnostic logs indicates the device entered backup ventilation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
EI 
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
EI  
Manufacturer Contact
kelly adams
2101 faraday ave
carlsbad, CA 92008
7606035046
MDR Report Key13929845
Report Number8020893-2022-00057
Device Sequence Number11203526
Product Code CBK
UDI-Device Identifier10884521592926
UDI-Public10884521592926
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K131252
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980X3ENDIUU-RMF
Device Catalogue Number980X3ENDIUU-RMF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Initial Report FDA Received Date03/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-