SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ENDOTRACHEAL TUBE; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/199/050
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when the customer put air in the cuff during the pre-use check, more resistance than usual was felt through the product.There was no patient injury reported.
 
Additional Manufacturer Narrative
One device was received for evaluation in relation to the reported event.Evaluation of the device is currently in progress.Once the evaluation is complete, a follow-up report with evaluation results will be submitted.
 
Additional Manufacturer Narrative
Other, other text: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample and two (2) photos were received for evaluation.Visual and functional testing were performed.Visual inspection of the photos found line occluded was detected in photo 2.Functional testing found inflation line occluded was observed, the complaint was confirmed.The root cause of the reported issue was found to be inflation line assembly not followed properly, therefore an occlusion was caused per excess solvent.Awareness notification was made to production personnel to explain the importance to adherence or following in the procedure on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ENDOTRACHEAL TUBE
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
3-10-15, hakata station east,
minneapolis, MN 55442
MDR Report Key13930697
Report Number3012307300-2022-05400
Device Sequence Number4003152
Product Code BTR
UDI-Device Identifier15019315022289
UDI-Public15019315022289
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K790312
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2025
Device Model Number100/199/050
Device Catalogue Number100/199/050
Device Lot Number3981457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2022
Supplement Date Received by Manufacturer08/05/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-