| Model Number |
1061T |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.The complaint includes blower and contaminated foam need to be replaced.There was no report of serious or permanent patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported that required intervention to prevent permanent impairment/damage, disability or permanent damage occured in section b2.After review, it was determined that all these were incorrectly selected in section b2.Nothing such has occured to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to the manufacturer's service center for evaluation.The manufacturer visually inspected the device and found water ingress and foam degradation in the unit.The manufacturer concludes there was evidence of sound abatement foam degradation and water ingress in the device.Section d2, g4 and h6 were updated/corrected in this report.
|
| |
|