Model Number SMXP130RUH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation could be performed.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) black female patient underwent a primary breast reconstruction surgery with implantation of an undisclosed breast tissue expander.Post-operatively, the patient was diagnosed with a deflated right tissue expander after an in-office visit with a medical professional.As a result, the patient underwent removal and replacement with a 560cc mentor memorygel breast implant on (b)(6), 2022.After removal the surgeon noted a tear where the buffer zone is.
 
Additional Manufacturer Narrative
On april 26, 2022, the mentor analysis lab received the suspect medical device for evaluation.On the device the lot number was provided and as such the identity of the device was provided.Device: 650cc artoura breast tissue expander: catalog: smxp130ruh, lot: 9431858.An evaluation was completed on april 27, 2022.Mentor then conducted a visual inspection, leak test and microscopic examination of the device.Visual analysis of the returned sample revealed that no damage was observed on the tissue expander.Leak testing was performed in accordance with mentor procedures, and a tear was observed at the juncture on the anterior patch.Additionally, no other leaks sites were observed.A microscopic examination was performed and the cause of the deflation could not be identified.As the authorization form for examination was not received the evaluation was limited to non-destructive testing; therefore, the shell thickness was not measured.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: causes of deflation of tissue expanders include but are not limited to the following events: fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.With the provided lot number, a manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key13932472
Report Number1645337-2022-03405
Device Sequence Number11929265
Product Code LCJ
UDI-Device Identifier00081317024392
UDI-Public00081317024392
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2024
Device Model NumberSMXP130RUH
Device Catalogue NumberSMXP130RUH
Device Lot Number9431858
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-