| Model Number |
H7493941824300 |
| Medical Device Problem Codes |
Break (1069); Material Rupture (1546); Activation Failure (3270)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that balloon rupture, partial deployment, and shaft break occurred.The target lesion was located in the obtuse marginal branch.The 3.00 x 24mm synergy xd stent was advanced for treatment.Upon inflation, the stent balloon ruptured leaving the stent partially deployed.When an attempt was made to remove the device, the catheter broke leaving the balloon and about 20cm of the shaft inside the patient.Wire placement was also lost.The patient was sent for emergency coronary artery bypass graft and the device portion retained in the patient was partially removed during surgery.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture, partial deployment, and shaft break occurred.The target lesion was located in the obtuse marginal branch.The 3.00 x 24mm synergy xd stent was advanced for treatment.Upon inflation, the stent balloon ruptured leaving the stent partially deployed.When an attempt was made to remove the device, the catheter broke leaving the balloon and about 20cm of the shaft inside the patient.Wire placement was also lost.The patient was sent for emergency coronary artery bypass graft and the device portion retained in the patient was partially removed during surgery.It was further reported via medwatch 2601800000-2022-8003 that the patient presented for stent placement in 3 locations.On the 3rd stent placement, the balloon appeared to have ruptured.A cardiothoracic surgeon contacted and the patient taken to surgery.The catheter and ruptured balloon were removed from the vessel.The surgeon and procedural cardiologist examined the catheter and balloon and determined them to be intact.The patient taken to cardiovascular intensive care unit for recovery and was discharged home.
|
| |
|
Search Alerts/Recalls
|