| Model Number |
IPN055190 |
| Medical Device Problem Codes |
Unraveled Material (1664); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "during use of the device we observed the guide was unraveled at its tip side, which made the remove of the needle difficult".No patient harm or consequence.It was reported the catheter was removed "with difficulties" and the doctor inserted a new catheter.
|
| |
|
Event or Problem Description
|
|
It was reported "during use of the device we observed the guide was unraveled at its tip side, which made the remove of the needle difficult".No patient harm or consequence.It was reported the catheter was removed "with difficulties" and the doctor inserted a new catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The customer returned one guide wire assembly and a lidstock for analysis.Signs of use in the form of biological material was observed on the guide wire body.The introducer needle was not returned for evaluation.Visual analysis revealed that the guide wire was severely unraveled from the distal end.One major kink was also observed.Microscopic examination confirmed the damage and revealed that the core wire separated directly adjacent to the distal weld.Visual analysis could not be performed on the introducer needle as it was not returned for analysis.The major kink on the guide wire measured 471mm from the proximal weld.The guide wire total length from the proximal weld to the point of separation on the core wire measured 602mm which is within the specification limits of 596mm-604mm per the guide wire product drawing.The guide wire outer measured outer diameter measured.790mm which is within the specification limits of.788mm-.826mm per the guide wire graphic.Dimensional analysis could not be performed on the introducer needle as it was not returned for analysis.The proximal end (functional end) of the guide wire was inserted through a lab inventory ars/introducer needle subassembly.No resistance was encountered as the guide wire passed completely through the assembly.Performed per ifu statement "advance guidewire into arrow raulerson syringe approximately 10 cm until it passes through syringe valves or into introducer needle".Functional analysis could not be performed on the introducer needle as it was not returned for analysis.A manual tug test confirmed that the proximal weld was secure and intact.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit informs the user, "do not withdraw guidewire against needle bevel to reduce risk of possible severing or damaging of guidewire".The ifu also states, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested." the report of an unraveled guide wire was confirmed through complaint investigation.Visual analysis revealed that the swg core wire had separated directly adjacent to the distal weld, which caused the component to unravel.The guide wire met all functional/dimensional requirements during investigation testing.Arrow guide wires of this size are designed and manufactured to withstand a tensile force of 2.75 pounds force.This internal specification is higher than the bs en iso 11070:2014 standard of 2.2 pounds force for this size wire.The selected insertion site and patient anatomy may present a tortuous path that could contribute to the possibility of guide wire kinking.Guide wire breakage may occur if a force greater than the design specification is applied during removal.Despite this, without the actual introducer needle returned for analysis, the root cause for this issue cannot be officially confirmed.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|