| Model Number |
CAG400S12 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being submitted as part of a batch submission of complaints that were reassessed as reportable foam degradation complaints; discovered as part of a retrospective remediation review.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information alleging an issue related to the cpap device's sound abatement foam.The device was returned, during device evaluation black soot was observed in device's top and bottom enclosures and blower box.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported incorrect information alleging an issue related to the cpap device's sound abatement foam.The device was returned, during device evaluation black soot was observed in device's top and bottom enclosures and blower box.There was no report of patient harm or injury.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.Section b5 should have been reported as: a cpap was returned to the manufacturer for service alleging black soot coming from the filter.There was no patient harm or injury reported.During the evaluation of the device at the manufacturer's service center, black soot contamination was observed in the top and bottom enclosures, blower and blower box.The unit was scrapped.Section e.1 was incorrectly reported in previous report, which is now updated in this follow-up report.
|
| |
|
Search Alerts/Recalls
|