| Model Number |
10005877 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the surgeon connected the mini cuff to apical holder and noted a plastic ring out of place.A new holder would be obtained and apical holder was removed.The patient had a stroke on (b)(6) 2022 following left ventricular assist device (lvad) implant.There were pupil changes but no signs of stroke showed on a computed tomography (ct) scan.A subsequent ct on (b)(6) 2022 showed development of ischemic changes in the right frontal lobe in the anterior aspect of the middle cerebral artery (mca) distribution.The patient was in the intensive care unit for monitoring and a neurology consult was scheduled.The patients care was withdrawn on (b)(6) 2022 due to non-recoverable stroke.The patient was on impella at p7 before the vad was implanted.The death was not considered to be device related and the device operated as intended.Related manufacturer reference report # 2916596-2022-01808.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation conclusion: the report of the o-ring becoming displaced from the apical cuff holding tool was confirmed through the submitted photo.A specific cause for this issue could not be conclusively determined through this evaluation.Review of the device history records (dhr) for the mini apical cuff kit, lot number 8182227,showed no deviations from manufacturing or quality assurance specifications.Review of the dhrs for the apical cuff holding tool assembly associated with this kit also showed no deviations from manufacturing or quality assurance specifications.Shop order found that the steps for confirming orientation on the sealing ring attachment were successfully performed for each holding tool within lot number 8126333.The heartmate 3 (hm3) mini apical cuff kit instructions for use (ifu) is currently available.The overview section of this ifu provides a description of the apical cuff holder, stating that a silicone o-ring is present at the interface of the circular disk and apical cuff, for a hemostatic connection with the apical cuff.This section also cautions the user to notify the appropriate personnel if there is a change in how the device works, sounds, or feels.Adverse events associated with the use of the mini apical cuff kit are also outlined; however, it should be noted that the patient did not experience any adverse consequences related to the current event.The surgical procedures section of the hm3 mini apical cuff kit ifu explains how to prepare the mini apical cuff and the apical cuff holder for use.No further information was provided.The manufacturer is closing the file on this event.
|
| |
|
Search Alerts/Recalls
|