BOSTON SCIENTIFIC CORPORATION WALLFLEX COLONIC; STENT, COLONIC, METALIC, EXPANDABLE Back to Search Results
Model Number M00565120
Medical Device Problem Codes Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Impact code: (b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
Note: this report pertains to one of two devices used during the same procedure.Refer to manufacturer report # 3005099803-2022-01697 for the associated device information.It was reported to boston scientific corporation on march 03, 2022 that a wallflex colonic stent was to be implanted in the transverse colon at the hepatic flexure as a bridge to surgery during a colonoscopy procedure performed on (b)(6) 2022.The patient's anatomy was tortuous and the patient was suspected with colon cancer.During the procedure, resistance was felt on the stainless steel handle of the device and the stent was unable to deploy (the subject of this report).While continuing to attempt to deploy the stent, the stainless steel handle bent.The stent was partially deployed when it was removed from the patient.A 90mm wallflex colonic stent (the subject mfr.# 3005099803-2022-01697) was attempted to be implanted; however, the stent fully deployed but remained partially inside the scope.Consequently, the stent was removed together with the scope.The procedure was not completed as another stent of the same size was unavailable.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX COLONIC
Common Device Name
STENT, COLONIC, METALIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13934292
Report Number3005099803-2022-01524
Device Sequence Number4731843
Product Code MQR
UDI-Device Identifier08714729456599
UDI-Public08714729456599
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K061877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/08/2023
Device Model NumberM00565120
Device Catalogue Number6512
Device Lot Number0027626975
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/03/2022
Initial Report FDA Received Date03/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-