• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC SAFE-T-J AMPLATZ EXTRA-STIFF SUPPORT WIRE GUIDE; DQX WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC SAFE-T-J AMPLATZ EXTRA-STIFF SUPPORT WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Model Number G03460
Medical Device Problem Code Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported, when preparing the device for a transcatheter aortic valve replacement, the inner core of a safe-t-j amplatz extra-stiff support wire guide protruded.The device did not make patient contact.Another device from the same product lot was used to complete the procedure.No adverse events to the patient were reported as a result of this occurrence.
 
Additional Manufacturer Narrative
Pma/510(k) #- k171764.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Additional Manufacturer Narrative
Event summary: as reported, when preparing the device for a transcatheter aortic valve replacement, the inner core of a safe-t-j amplatz extra-stiff support wire guide protruded.The device did not make patient contact.Another device from the same product lot was used to complete the procedure.No adverse events to the patient were reported as a result of this occurrence.Investigation - evaluation: reviews of the complaint history, device history record, instructions for use, manufacturing instructions, and quality control procedures and a visual inspection of the device were conducted during the investigation.One safe-t-j amplatz extra-stiff support wire guide (thscf-35-260-3-aes) was returned to cook for investigation in a prior to use condition.Approximately 1cm from the distal tip of the wire guide, and the wire was exposed.A document-based investigation evaluation was performed.No related non-conformances were recorded, and there have been no other reported complaints for this lot number.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which caution, ¿inspect the wire guide for kinks or damage prior to use.¿ based on the available information, cook has concluded that shipping/handling contributed to this incident.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE-T-J AMPLATZ EXTRA-STIFF SUPPORT WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key13934367
Report Number1820334-2022-00473
Device Sequence Number4886730
Product Code DQX
UDI-Device Identifier00827002034600
UDI-Public(01)00827002034600(17)270110(10)14427779
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG03460
Device Catalogue NumberTHSCF-35-260-3-AES
Device Lot Number14427779
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date05/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-