| Model Number |
M00561191 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).(investigation result): one captivator snare was received for analysis.Visual analysis and microscopic evaluation of the returned device revealed that the 2 in 1 (cautery pin) was broken inside of the active cord.Functional evaluation of the returned device was impossible due the device condition.The reported event of "device difficult to actuate" was confirmed since it's probable that the broken 2 in 1 (cautery pin) induced difficulty in actuating the device.Based on the analysis of the returned device and the information available, the code selected as the most probable cause is design inadequate for purpose.There is an investigation in place to address this problem.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a 25mm captivator ii round stiff snare was used during a procedure performed on (b)(6) 2022.During the procedure, the device was difficult to actuate.The procedure was completed with a different device.There were no patient complications reported as a result of this event.Note: this event has been deemed an mdr-reportable event based on investigation results which revealed that the cautery pin was detached.Please see block for full investigation details.
|
| |
|
Search Alerts/Recalls
|