BIOMET MICROFIXATION 1.5 MM SYSTEM TWIST DRILL .8X50MM 7MM STOP 1.0; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the drill tip fractured while drilling the pilot hole and about a 5mm drill tip remains in the body.The surgeon plans to remove the drill tip at least a month after the bone has healed.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A visual inspection was conducted on the returned twist drill.The drill shows sings of use including marking on the drill body.The drill has fractured where the drill flutes meet the drill base.The complaint is confirmed.A determination cannot be made as to what caused the drill to fracture.Review of the cert identified no deviations or anomalies during manufacturing.The supplier dhr was not requested because the raw material certificate showed material is conforming to specification and no problems were found after the return of the devices.The dhr was not reviewed as zimmer lot is for packaging purposes only and the complaint is not related to a packaging related issue.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: metallic plate fixation of the left orbital floor is noted and a small drill tip fragment is present at the medial orbital floor.Additional plates and screws are present.No displacement of the fractures is seen and alignment of the post-operative facial bones appears anatomic.Bone quality appears unremarkable.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.This report is being submitted to update additional information in section b4, b5, d9, g3, g6, h2, h3, h4, h6 and h10.Correction is in section d4.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5 MM SYSTEM TWIST DRILL .8X50MM 7MM STOP 1.0
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
christina arnt
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key13934487
Report Number0001032347-2022-00099
Device Sequence Number14635634
Product Code HBE
UDI-Device Identifier00841036014247
UDI-Public(01)00841036014247(10)318282
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062842
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number01-7428
Device Lot Number318282
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2022
Supplement Date Received by Manufacturer04/11/2022
Initial Report FDA Received Date03/28/2022
Supplement Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/13/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-