• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. PLATE HOLDER, THORACOLUMBAR; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. PLATE HOLDER, THORACOLUMBAR; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 1208-90007
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
A company representative reported that a cayman plate holder was not holding the plate "properly." the procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Event or Problem Description
A company representative reported that a cayman plate holder was not holding the plate "properly." the procedure was completed successfully with no surgical delay and no adverse consequence to the patient.
 
Additional Manufacturer Narrative
A visual inspection was performed and there appeared to be no notable damage or deformation on the tip that would have caused the event.The device was functionally tested with a sample plate and it was able to release and hold the implant without any mechanical issues arising.As the device was able to perform its intended function the reported event was unable to be replicated.Device history records were reviewed for the corresponding lot and no relevant manufacturing issues were identified.A review of complaint history associated with the subject lot and catalog number was performed and no adverse trends were observed.Limited information regarding the event was made available to stryker.Multiple attempts were made to obtain additional information, but no response was received.Based on the inspection of the device it is most likely that the surgeon in this case misassembled the device during surgery which led to the device not holding the plate properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATE HOLDER, THORACOLUMBAR
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13934721
Report Number3004774118-2022-00115
Device Sequence Number20055299
Product Code LXH
UDI-Device Identifier10888857019645
UDI-Public10888857019645
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1208-90007
Device Lot NumberAEET
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/01/2022
Supplement Date Received by Manufacturer05/26/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/05/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-