MICROVENTION, INC. MICROPLEX COSMOS; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 182065CS-V
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformance's associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was returned to the manufacturer for evaluation.The investigation is currently underway.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
 
Event or Problem Description
It was reported that an embolization implant coil detached unexpectedly.The implant was removed without the need for intervention.There was no reported injury to the patient.The patient's condition was reported to be "good.".
 
Additional Manufacturer Narrative
The investigation of the returned coil system found the implant intact with no notable damage and to be attached to the pusher.No burn marks were found on the pusher's heater coil.The investigation found the pusher's body coil damaged 2" from the distal heater coil and lead wire separation starting 24" from the proximal connector.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.The investigation of the returned device did not find any damage or other anomaly that would have caused or contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROPLEX COSMOS
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terry callahan
35 enterprise drive
aliso viejo, CA 92656
7142478000
MDR Report Key13935101
Report Number2032493-2022-00113
Device Sequence Number4132162
Product Code HCG
UDI-Device Identifier00810170014185
UDI-Public(01)00810170014185(11)200311(17)250228(10)2003115Q1
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
K090891
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model Number182065CS-V
Device Catalogue Number182065CS-V
Device Lot Number2003115Q1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2022
Supplement Date Received by Manufacturer03/31/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date04/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age40 YR
Patient SexMale
Patient Weight74 KG
-
-