| Model Number |
110-150 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Swollen Lymph Nodes/Glands (4432)
|
| Date of Event |
12/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side biocell replacement, "multiple enlarged lymph nodes in right axilla.", and rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.The event of "lymphadenopathy" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: underweight, an opening on posterior, crease fold, and wear abrasion.A microscopic analysis was performed which identified: a crease sharp opening on posterior.Based on the device analysis the final assessment is: a crease sharp opening on posterior assessed as fold flaw opening.Creases/folding of implant condition was observed, nevertheless it is not related to the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side biocell replacement, "multiple enlarged lymph nodes in right axilla.", and rupture.Device has been explanted.
|
| |
|