| Model Number |
LSMU0800858 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device-Device Incompatibility (2919); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 10/2024).
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure of a highly calcified target lesion in the bilateral common iliac artery via a bilateral femoral approach, the device was allegedly unable to reach the lesion.The stent was reported to be stuck with the tip of the sheath upon attempting to exchange the guidewire, which led to a stent dislodgement from the delivery system.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported for this product/lot number combination.Therefore, a device history record review is not required.Investigation summary: the sample was not returned for evaluation.The result of the investigation is inconclusive for the reported stent dislodgment and failure to advance issues.The definitive root cause could not be determined based upon available information.Labeling review: the instructions for use for the lifestream product was reviewed and contains the following information relevant to the reported event: the lifestream balloon expandable vascular covered stent endovascular system is available in catheter lengths of 80 cm and 135 cm and it is compatible with 0.035" (0.89 mm) guidewires.The premounted covered stent is available in multiple diameters and lengths.Warnings: should excessive resistance be felt at any time during the insertion process, do not force passage.Attempts to retract the covered stent into the sheath/guiding catheter may result in dislodgement and embolization of the covered stent.Maintain guidewire placement across the lesion and withdraw the endovascular system only until the proximal end of the covered stent is aligned with the tip of the sheath/guiding catheter.Do not attempt to remove an unexpanded covered stent through the sheath/guiding catheter.Remove the sheath/guiding catheter and endovascular system as a single unit.Attempting to remove an unexpanded covered stent by pulling it back into the sheath/guiding catheter may result in stent dislodgement.The covered stent cannot be repositioned after it is deployed.Precautions: the device should only be used by physicians who are trained in endovascular procedures and are familiar with the complications, side effects and hazards of peripheral vascular interventions.Prior to device use, refer to the covered stent sizing table on the label and read the instructions for use.Crossing the implant with catheters or other adjunct devices can result in covered stent dislodgement or damage.Potential patient/device adverse effects that may occur include, but are not limited to, the following: covered stent dislodgement from balloon during tracking procedure.Covered stent misplacement during placement procedure.Directions for use: site access and preparation.1.Using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.2.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Covered stent size selection 3.Select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation 4.Carefully remove the selected device from the package.5.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.Introduction of the endovascular system and placement of the covered stent 13.Advance the endovascular system over the guidewire into the introducer sheath.14.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.15.Slowly inflate the endovascular system balloon to nominal pressure, expanding the covered stent.Confirm complete expansion via fluoroscopic visualization.A 15 ¿ 30 second inflation time is recommended.H10: d4 (expiry date: 10/2024).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure of a highly calcified target lesion in the bilateral common iliac artery via a bilateral femoral approach, the device was allegedly unable to reach the lesion.The tip of the sheath collided with the stent when attempting to exchange the guidewire, which led to a stent dislodgement from the delivery system.The procedure was completed using another device.There was no reported patient injury.
|
| |
|