• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S6; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S6; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5536-B-600
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This is a product mix issue, so the device information located on the outer box label is being reported in section d and the information for the device that was actually inside the package is : cat# titanium bplate s5; cat # 5536b500; lot # ctd69106 product field action has been initiated.Investigation ongoing.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Primary procedure, left mako tka.It was reported that the packaging and labels were for a size 6 press-fit baseplate (catalog and lot).The implant was a size 5 press-fit baseplate (different catalog and lot).Discovery was made when the surgeon attempted to implant a size 6 insert and it wouldn't fit.Hospital staff then called out the catalog and lot on the implant to the rep.Surgeon elected to leave the size 5 baseplate implanted and implanted a fitting insert.Surgery completed successfully with a delay of approximately 5 minutes.Rep reported he can provide pictures of the packaging (but not of the implant as it was in the patient/ sterile field by that time).Rep also confirmed that no further information will be released by the hospital or surgeon.
 
Additional Manufacturer Narrative
Reported event: an event regarding product mix involving a triathlon baseplate was reported.The event was confirmed through evaluation of the provided photographs.Method & results: product evaluation and results: the device itself was not returned as it was implanted however photographs of the packaging were provided.The photographs show the unit carton, labels containing the catalog and lot details for a size 6 baseplate are fixed to the unit carton.Writing with black pen on the label states that inside the box was a size 5 baseplate.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: visual inspection of the returned packaging confirmed the event.Nc was raised to address this event.
 
Event or Problem Description
Primary procedure, left mako tka.It was reported that the packaging and labels were for a size 6 press-fit baseplate (catalog and lot).The implant was a size 5 press-fit baseplate (different catalog and lot).Discovery was made when the surgeon attempted to implant a size 6 insert and it wouldn't fit.Hospital staff then called out the catalog and lot on the implant to the rep.Surgeon elected to leave the size 5 baseplate implanted and implanted a fitting insert.Surgery completed successfully with a delay of approximately 5 minutes.Rep reported he can provide pictures of the packaging (but not of the implant as it was in the patient/ sterile field by that time).Rep also confirmed that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRITANIUM BPLATE TRIATHLON S6
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13937739
Report Number0002249697-2022-00447
Device Sequence Number11881309
Product Code MBH
UDI-Device Identifier07613327041514
UDI-Public07613327041514
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5536-B-600
Device Catalogue Number5536B600
Device Lot NumberCTD69463
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/03/2022
Supplement Date Received by Manufacturer06/15/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date07/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1237-2022
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age74 YR
Patient SexMale
-
-