This report is being submitted to report additional information.The following sections are being reported: d4: expiration date and unique identifier (udi) number h2: if follow-up, what type h3: device evaluated by manufacturer h4: device manufacturer date h6: type of investigation h6: investigation findings h6: investigation conclusions h10: additional narratives/data the product was returned but the reported event could not be verified as the exact details of event and patient anatomical conditions were unknown/nonverifiable.Based on the evaluation and dimensional analysis, device malfunction has not occurred.Additionally, there is no existing actionable trend or non-conformance / capa / hhe / d / ie / product hold against the reported device that did or could cause or contribute to the reported event.Monthly post market trending review identified no actionable trends or corrective actions for the reported event or device.Zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming products.Therefore, based on the available information, the product was within specifications and conforming when it left zimmer biomet.Dhr review for the lot had revealed no deviations nor non-conformances which could have caused or contributed to the reported event.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The device could not be functionally tested for the reported failure/event.Therefore, the reported event could not be recreated.The complaint is not related to the functional performance of the device.A definitive root cause could not be determined.No further investigation or immediate capa/hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|