| Model Number |
G48038 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # k163468.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
User advanced the delivery system to stenosis, but found out the stent shape deformed during deployment.See attached pic.User retracted the device from patient, then changed to another same device to complete the procedure.((b)(4)).Lab evaluation date: 02 mar 2022: visual inspection: lock wire not returned.Red marker at 4th dimpling on the handle.Stent wire appears to be exposed on return.Functional inspection: handle actuated as intended.Stent deployed without issue but appears to be visually deformed.This file will capture user error, evo-c is not intended to be removed or repositioned after stent placement.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|