• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD EVOLUTION® COLONIC CONTROLLED-RELEASE STENT - UNCOVERED; MQR STENT, COLONIC METALLIC EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD EVOLUTION® COLONIC CONTROLLED-RELEASE STENT - UNCOVERED; MQR STENT, COLONIC METALLIC EXPANDABLE Back to Search Results
Model Number G48038
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) # k163468.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
User advanced the delivery system to stenosis, but found out the stent shape deformed during deployment.See attached pic.User retracted the device from patient, then changed to another same device to complete the procedure.((b)(4)).Lab evaluation date: 02 mar 2022: visual inspection: lock wire not returned.Red marker at 4th dimpling on the handle.Stent wire appears to be exposed on return.Functional inspection: handle actuated as intended.Stent deployed without issue but appears to be visually deformed.This file will capture user error, evo-c is not intended to be removed or repositioned after stent placement.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION® COLONIC CONTROLLED-RELEASE STENT - UNCOVERED
Common Device Name
MQR STENT, COLONIC METALLIC EXPANDABLE
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
aisling hassett
o halloran road
national technology park
limerick 
MDR Report Key13939063
Report Number3001845648-2022-00183
Device Sequence Number11881324
Product Code MQR
UDI-Device Identifier10827002480381
UDI-Public(01)10827002480381(17)230112(10)C1791590
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2023
Device Model NumberG48038
Device Catalogue NumberEVO-25-30-10-C
Device Lot NumberC1791590
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/29/2021
Event Location Hospital
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/10/2022
Initial Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-