• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HEM-O-LOK XL ENDO 10, APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HEM-O-LOK XL ENDO 10, APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Model Number IPN915193
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.
 
Event or Problem Description
When the user ligated a renal artery with a clip during a laparoscopic nephrectomy , the clip got stuck in the applier.The user touched the clip carefully with a forceps, then the clip left the applier.The clip did not involve other tissue at ligation.Additional information: the clip got stuck in the applier after ligation, therefore, when the clip left the applier , ligation succeeded.
 
Additional Manufacturer Narrative
(b)(4).Per customer provided information we are unable to perform a thorough dhr review for the alleged defective device since lot information has not been provided and this instrument has not been returned for review or evaluation.We are unable to determine what may have caused the alleged defect or validate the alleged complaint.All devices of this type are 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Event or Problem Description
When the user ligated a renal artery with a clip during a laparoscopic nephrectomy , the clip got stuck in the applier.The user touched the clip carefully with a forceps, then the clip left the applier.The clip did not involve other tissue at ligation.Additional information: the clip got stuck in the applier after ligation, therefore, when the clip left the applier , ligation succeeded.
 
Additional Manufacturer Narrative
(b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet, inc.Kenosha wi facility as part of a 50-piece lot in january of 2019.Evaluation of the returned instrument shows that there is no visible damage to the jaws, and they are properly aligned in the open and closed position and the open and closed jaw gaps measure to print specifications of.526 +.010/-.020 at (.512) and.004/.012 at (.008).We are unable to find any non-conforming features on this device.Functional evaluation shows that this instrument is able to pick-up, retain, close and release multiple clips both with and without the use of silastic test tubing as it did at time of manufacture.We are unable to validate this complaint as we were unable to replicate the alleged issue at the end user's facility.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Event or Problem Description
When the user ligated a renal artery with a clip during a laparoscopic nephrectomy, the clip got stuck in the applier.The user touched the clip carefully with a forceps, then the clip left the applier.The clip did not involve other tissue at ligation.Additional information: the clip got stuck in the applier after ligation, therefore, when the clip left the applier, ligation succeeded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEM-O-LOK XL ENDO 10, APPLIER
Common Device Name
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key13939087
Report Number3011137372-2022-00056
Device Sequence Number4309221
Product Code GDO
UDI-Device Identifier24026704696820
UDI-Public24026704696820
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN915193
Device Catalogue Number544990
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer05/09/2022
05/23/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date05/09/2022
05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-