| Catalog Number |
106696 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.However, based on the information provided, there was a crack on the deaeration chamber of the return line and the most probable cause was identified as a shock on the device during transportation.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during priming with one unit of prismaflex m60 set, a crack was observed in the de-aeration chamber and fluid leaked out resulting in an unspecified alarm being triggered.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to d9, h3, h6 and h10.H0: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was received for evaluation.A pressure leak test was performed and a leak was observed at the level of the degassing chamber (at the weld of the chamber assembly).The reported condition was verified.The most probable cause was identified as a shock on the device during shipping.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|