MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS M; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX SETS M; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 106696
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.However, based on the information provided, there was a crack on the deaeration chamber of the return line and the most probable cause was identified as a shock on the device during transportation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that during priming with one unit of prismaflex m60 set, a crack was observed in the de-aeration chamber and fluid leaked out resulting in an unspecified alarm being triggered.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Additional information was added to d9, h3, h6 and h10.H0: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was received for evaluation.A pressure leak test was performed and a leak was observed at the level of the degassing chamber (at the weld of the chamber assembly).The reported condition was verified.The most probable cause was identified as a shock on the device during shipping.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SETS M
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13939265
Report Number8010182-2022-00080
Device Sequence Number6505380
Product Code KDI
UDI-Device Identifier07332414064549
UDI-Public(01)07332414064549
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K041005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Catalogue Number106696
Device Lot Number21F0809
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer04/25/2022
Initial Report FDA Received Date03/29/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NA.
Patient SexUnknown
-
-