| Model Number |
4351-35 |
| Medical Device Problem Codes |
Extracardiac stimulation (1463); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Electric Shock (2554)
|
| Date of Event |
03/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that the patient of the state is experiencing shock at the pocket site.Needs assistance in turning off therapy or turning amplitude down.The caller was not with the patient.The caller was redirected to warm transferred to nas to page a local rep.No further complications were reported/anticipated at this time.
|
| |
|
Event or Problem Description
|
|
Additional information was received from a healthcare professional (hcp).The hcp reported that the steps taken to resolve the reported issue was to interrogate the device.It was noted the battery was at eol (end of life) and the battery was not functioning when this occurred.The issue was resolved at the time of the report.The device is still implanted /off/eol.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|