Device was returned for analysis on 03/07/2022.Aspect control : the device returned in dry conditions in its original pack.Estimated insertion depth 55mm (tightening marks).The device presents few markings and a yellow color to an orange color on the pa tube.Functional control : when first connected (in return state), the displayed values of temperature and pressure are stable.When simulated in testing, the sensor reacts normally.When tested in a water column under different pressure values, the sensor is stable, for 24 hours.Conclusion : in its returned state, the device is within specifications and the reported problem cannot be reproduced nor confirmed.
|