• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOPHYSA SOPHYSA; PRESSIO® ICP & ICT MONITORING KIT, PARENCHYMAL WITH BOLT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOPHYSA SOPHYSA; PRESSIO® ICP & ICT MONITORING KIT, PARENCHYMAL WITH BOLT Back to Search Results
Model Number PSO-PBT
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was returned for analysis on 03/07/2022.Aspect control : the device returned in dry conditions in its original pack.Estimated insertion depth 45mm (tightening marks).The device presents few markings and a yellow color to an orange color on the pa tube.Functional control : when first connected (in return state), the displayed values of temperature and pressure are stable.When simulated in testing, the sensor reacts normally.When tested in a water column under different pressure values, the sensor is stable, for 24hours.Conclusion : in its returned state, the device is within specifications and the reported problem cannot be reproduced nor confirmed.
 
Event or Problem Description
The physician reported a difference between icp values and displayed amplitude on the monitor; no health consequences reported on the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOPHYSA
Common Device Name
PRESSIO® ICP & ICT MONITORING KIT, PARENCHYMAL WITH BOLT
Manufacturer (Section D)
SOPHYSA
5 rue guy moquet
orsay, 91400
FR  91400
Manufacturer (Section G)
SOPHYSA
temis
rue sophie germain
besançon, 25000
FR   25000
Manufacturer Contact
marion prudhomme
temis
rue sophie germain
besançon, 25000
FR   25000
MDR Report Key13939420
Report Number3001587388-2022-22090
Device Sequence Number4792007
Product Code GWM
UDI-Device Identifier03760124131758
UDI-Public(01)03760124131758(11)210818(17)260801
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K162108
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2026
Device Model NumberPSO-PBT
Device Catalogue NumberPSO-PBT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/28/2022
Initial Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age45 YR
Patient SexFemale
-
-