MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M Back to Search Results
Model Number BI70000028120
Medical Device Problem Code Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the issue could be recreated.It was verified that the issue was not the umbilical cable.After reseating all connectors on pleora and pax scan, the issue disappeared.The rep took many x-ray shots and performed several reboots.The issue appeared to be resolved by reseating cables.The imaging system passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the imaging acquisition system (ias) would occasionally switch into stand-alone mode when in use.The issue discovered outside of case and there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Common Device Name
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key13939424
Report Number3004785967-2022-00223
Device Sequence Number4792009
Product Code OXO
UDI-Device Identifier00643169845091
UDI-Public00643169845091
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K092564
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBI70000028120
Device Catalogue NumberBI70000028120
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Initial Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-