Model Number PSO-PBT
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
The physician reported a difference between icp values and displayed amplitude on the monitor; no health consequences reported on the patient.
 
Additional Manufacturer Narrative
Device was returned for analysis on 03/07/2022.Aspect control: the device returned in dry conditions in its original pack.Presence of blood marks on cone and pa tube.The rest of the device was relatively clean.Functional control : when first connected (in return state), the displayed values of temperature and pressure are stable.When simulated in testing, the sensor reacts normally.When tested in a water column under different pressure values, the sensor is stable, for 24 hours.Conclusion: in its returned state, the device is within specifications and the reported problem cannot be reproduced nor confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOPHYSA
Common Device Name
PRESSIO® ICP & ICT MONITORING KIT, PARENCHYMAL WITH BOLT
Manufacturer (Section D)
SOPHYSA
5 rue guy moquet
orsay, 91400
FR  91400
Manufacturer (Section G)
SOPHYSA
temis
rue sophie germain
besançon, 25000
FR   25000
Manufacturer Contact
marion prudhomme
temis
rue sophie germain
besançon, 25000
FR   25000
MDR Report Key13939429
Report Number3001587388-2022-22092
Device Sequence Number14650643
Product Code GWM
UDI-Device Identifier03760124130768
UDI-Public(01)03760124130768(11)210817(17)260801
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K162108
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2026
Device Model NumberPSO-PBT
Device Catalogue NumberPSO-PBT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age75 YR
Patient SexFemale
-
-