Device was returned for analysis on 03/07/2022.Aspect control: the device returned in dry conditions in its original pack.Presence of blood marks on cone and pa tube.The rest of the device was relatively clean.Functional control : when first connected (in return state), the displayed values of temperature and pressure are stable.When simulated in testing, the sensor reacts normally.When tested in a water column under different pressure values, the sensor is stable, for 24 hours.Conclusion: in its returned state, the device is within specifications and the reported problem cannot be reproduced nor confirmed.
|