|
STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 32/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
|
Back to Search Results |
|
| Model Number |
6570-0-232 |
| Medical Device Problem Code |
Inaccurate Information (4051)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding product mix involving a ceramic head was reported.The event was confirmed by review of the photographs provided.Method & results: product evaluation and results: the outer box of the device shows catalog number 6570-0-232, lot 89546202 on the label and also shows that the device was intended to be a 32mm head with a nk lnth of +4mm.A photo showing that the device packaged bears the marking 32mm -4.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been 8 other similar events for the lot referenced.Conclusions: the reported product mix was confirmed.An nc was raised to further address this event.Product surveillance will continue to monitor for trends.Corrective action/preventive action: nc was raised to further address this event.This is a product mix issue, so the device information located on the outer box label is being reported and the information for the device that was actually inside the package is: 6570-0-032 delta v-40 ceramic head 32/-4 lot 89648802 product field action has been initiated.
|
| |
|
Event or Problem Description
|
|
Primary procedure, left hip.It was reported that an implant's packaging and labels indicated a 32 +4 ceramic head, but the device was marked as a 32 -4 ceramic head.A second device of the same lot was opened and the same issue occurred.A 32 +4 head of a different lot was opened and implanted.Procedure completed successfully with a delay of approximately 2 minutes.Rep confirmed the implants were hospital owned.Rep confirmed that he can provide pictures, and that no further information will be released by the hospital or surgeon.
|
| |
|
Search Alerts/Recalls
|
|
|